播散性隐球菌病患者的血清隐球菌抗原假阴性侧流免疫测定结果

 

Authors Xu Y, Xia W, Ni F

Received 2 June 2020

Accepted for publication 28 July 2020

Published 18 August 2020 Volume 2020:13 Pages 2877—2881

DOI https://doi.org/10.2147/IDR.S265784

Checked for plagiarism Yes

Review by Single anonymous peer review

Peer reviewer comments 2

Editor who approved publication: Professor Suresh Antony

Abstract: Cryptococcus neoformans (C. neoformans ) is an opportunistic fungal pathogen to humans, which can be acquired from environmental sources. Its most important virulence factor is its polysaccharide capsule, which can be used for diagnostic tests that identify the cryptococcal antigen (CrAg). The CrAg lateral flow assay (LFA) is a dipstick immunochromatographic assay with high sensitivity and specificity; however, several false-negative cases have been reported. Here, we present a case of a false-negative serum CrAg LFA, in which the blood culture from a matched sample was positive for C. neoformans, thus demonstrating the postzone phenomenon.
Keywords: Cryptococcus neoformans , lateral flow immunoassay, postzone phenomenon




Figure 1 (A) Morphology of C. neoformans in blood culture bottle by...