已发表论文

新型细胞形态分析仪在头颈癌中的应用

 

Authors Zhang Y , Xie M, Xue R, Tang Q, Zhu X, Wang J, Yang H, Ma C 

Received 26 September 2021

Accepted for publication 23 November 2021

Published 3 December 2021 Volume 2021:14 Pages 9307—9314

DOI https://doi.org/10.2147/IJGM.S341420

Checked for plagiarism Yes

Review by Single anonymous peer review

Peer reviewer comments 2

Editor who approved publication: Dr Scott Fraser

Purpose: Rapid and accurate diagnosis of the pathological characteristics of head and neck cancer and tumor resection margins is important. The DiveScope cell morphology analyzer (DiveScope) is a new endomicroscope that can rapidly image living tissues and cells. In this study, we preliminarily examined the accuracy of the DiveScope for determining the malignancy of head and neck cancers and the positivity/negativity of tumor resection margins and determined the consistency between diagnostic results with the DiveScope and those of frozen section pathology to provide a foundation for further clinical trials in pathological diagnosis of head and neck cancers and tumor resection margins.
Methods: Head and neck cancer samples and resection margin samples were rapidly stained ex vivo before observation under the DiveScope cell morphology analyzer. Experienced pathologists distinguished the benignity and malignancy of the tumors based on images obtained by the DiveScope in a double-blind manner to validate the diagnostic performance of the analyzer.
Results: We found that the cell morphology, cell nucleus morphology, karyoplasmic ratio, and even the nucleolus could be clearly distinguished. The positive likelihood ratio (PLR) and negative likelihood ratio (NLR) of benign and malignant head and neck cancer according to DiveScope results were 10.55 and 0.04, respectively. The PLR and NLR of the head and neck cancer resection margins according to the DiveScope were 19.01 and 0, respectively.
Conclusion: The DiveScope showed high accuracy in determining the benignity and malignancy of head and neck cancer and the positivity/negativity of resection margins, and its results were highly consistent with those of intraoperative frozen section pathology tests.
Keywords: the DiveScope cell morphology analyzer, head and neck cancer, resection margins, pathological diagnosis